Customization: | Available |
---|---|
Type: | Anesthesia Multigas Analyser |
OEM: | Acceptable |
Still deciding? Get samples of US$ 1395/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product name | Anesthesia multigas module |
Model | AG800 |
MOQ | 1 |
Transducer Type | UN-100 series multigas analyzer | ||
Sample Rate | 50mL/min,±10mL/min | ||
Principle of Operation | Non-dispersive infrared(NDIR),no moving part | ||
Initialization Time | 20 sec,full specifications within 60sec | ||
Calibration | No routine user calibration required | ||
Compensation | Automatic for atmospheric pressure,temperature | ||
Rise Time | CO2<200ms, N2O,AA<350ms | ||
Respiratory Rate | Range:3-150 BPM Accuracy ±1BPM |
||
Breath Detect | Adaptive threshold,minimum 1% CO2 change | ||
Agent Threshold | Agent 0.15vol% | ||
Measurement Accuracy | Gas Type | Range | Accuracy |
CO2 | 0-10% | ±(0.2vol%+2% of reading) | |
10-15% | ±(0.2vol%+2% of reading) | ||
N2O | 0-100% | ±(2vol%+2% of reading) | |
ISO | 0-6% | ±(0.2vol%+2% of reading) | |
ENF | 0-6% | ±(0.2vol%+2% of reading) | |
SEV | 0-8% | ±(0.2vol%+2% of reading) | |
Interface | RS-232 Operaing at 9600bps | ||
Voltage Requirements | 5.0DV±5% | ||
Power Consumption | Less than 2.0 Watts typical | ||
Temperature and Humidity | Operating:0-40ºC,10-90% RH,non-condensing Storage: -40-70ºC,<90%RH, non-condening |
||
Size | 96*57.5*30mm | ||
Weight | 110g |